Issue: Sun Angel S52 Duo displays E132 when "Essential Sunshine" is selected.

Article ID: 1081
Last updated: 28 Feb, 2013
Revision: 1
Add comment
Views: 0
Comments: 0

If the Sun Angel S52 Duo displays E132 when "Essential Sunshine" is selected it is likely that a wire on the K3 relay is in the incorrect position.  To resolve this issue remove the wire from K13 Position 13 and move to position 14 on the same relay. 

Note:  Modification of the wiring in any electrical device should only be performed by qualified persons. Please contact your local JK Distributor for service.

This article was:  
Prev   Next
General Information: Serial Number Locations on Ergoline and...     Sun Angel S52

RSS

Contraindication: This product is contraindicated for use on persons under the age of 18 years. "Contraindication: This product must not be used if skin lesions or open wounds are present." "Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer." "Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.